Volume 8, Issue 4, October 2018
Main Menu
SIMULTANEOUS ESTIMATION OF IMIDAPRIL HYDROCHLORIDE AND HYDROCHLOROTHIAZIDE BY RP-HPLC
Author(s): Veditha.K* and Sailaja.B
Abstract: A simple isocratic reverse phase high performance liquid chromatography (RP-HPLC) method has been developed
and subsequently validated for simultaneous estimation of anti-hypertensive drugs, Imidapril(IMP) and
Hydrochlorothiazide(HTZ) in tablet formulation. Quantification was carried out by using an isocratic mode on HPLC
of Phenomenex Luna (C18 250 X 4.6 mm, 5 microns) with flow rate of 1.0 ml/min and UV detection at 218nm. The
separation was carried out using a mobile phase consisting of phosphate buffer: acetonitrile: methanol (58: 25:
17 v/v) and adjusted the pH to 2.5. The retention time for IMP and HTZ was found to be 6.243min and 2.370min
respectively. A linear response was observed over the concentration range of 5-30 mcg/ml each for the assay of
IMP and HTZ. The results of analysis were validated and recovery was found to be 98%. Hence the proposed
method was found to be accurate, precise, reproducible and specific and can be used for simultaneous analysis of
these drugs in pharmaceutical formulation.
PAGES: 68-76 | 40 VIEWS 46 DOWNLOADS
How To Cite this Article:
Veditha.K* and Sailaja.B. SIMULTANEOUS ESTIMATION OF IMIDAPRIL HYDROCHLORIDE AND HYDROCHLOROTHIAZIDE BY RP-HPLC. 2018; 8(4): 68-76.